Evolve from the complicated to the simple! Simplicity means user-friendliness, wonderful handling, superb functionality, durability, best treatment results and true reliability
Read more: CataRhex Swisstech Concept
The peristaltic pump system of SwissTech beats its competitors! It achieves, without difficulty, the performance levels of a modern venturi system. The vacuum build only takes parts of a second. Whether it is to be vacuumised very finely or with full power, with the SwissTech peristaltic pump the flow is under precise pedal control at all times. Even with very low flow, no rolling effect is detectable.
The appliance of correct scientific principles and the elimination of air in the fluid system are key elements of the great chamber stability of the SwissTech. Therefore, high performance levels for flow and vacuum can be obtained without difficulty. SwissTech is ideal for the easyPhaco® technology!
The patent-registered pressure measuring principle prevents the sensor from contact with the aspiration fluid. The pressure sensor integrated into the tube system eliminates the risk of contamination and is totally free of air.
CataRhex® SwissTech allows for the most up-to-date operating techniques in cataract surgery:
|
VC820100SP CataRhex®SwissTech I/A, phaco (pulse, burst, continuous, easyPhaco®, MICS, CMP), anterior vitrectomy, capsulotomy, diathermia, glaucoma STT and IDK, mountable to all IV poles, peristaltic pump system, DirectAccess® control panel Scope of delivery: basic unit, control pedal, connection cable, instruction manual, IV pole or transport case, tubing set |
|
|
SwissTech Control Pedal Four areas of movement for a fine linear control of all instruments, retrieval of memory and switching on and off of functions. The pedal is waterproof, requires very little space, can be moved aside by foot and allows fatigue-free handling. |
|
|
VC830201 NovitreX® Vitrectomy Unit A compact unit that can be combined with SwissTech. With Goodlight® metal halide light source, air-fluid exchange, visco fluid injection and extraction, DirectAccess® control panel. |
|
|
VE820003 Transport Case (Pilot Suitcase) Makes possible easy transport as hand-luggage. Everything can be put in, including the SwissTech unit, control pedal, instruments and waste bottle. |
![]() |
I/A System
| Peristaltic: | 0 to 600 mmHg, 5 mmHg increments, 1 bis 50 ml/min, 1 mml/min increments, rise time < 0.6s (50 ml/600 mmHg) |
| Reflux: | from bottle, pedal deflection backwards or sidewards |
| Memories: | 2, proportional override function |
| Tubing System: | closed with integrated pressure sensor, single use or autoclavable |
Ultrasound Phaco
| Frequency: | 28 kHz, piezo driver |
| Control: | Autotuning (50 times/s), linear, pulsed, CMP cool micro pulse, burst, occlusion mode |
| Power: | 1 to 100% in steps of 1%, max. deflection 1.2 um |
| Burst: | duration 10 ms to 500 ms, pause 2s to continuous |
| Puls: | pulse frequency 0.5 to 40Hz, pulse duration 0.25 to 1800 ms |
| Memories: | 3 |
| Counter: | effective (relevant to energy delivered), 0.1s increments |
Bipolar Diathermy
| Frequency: | 500 kHz, pulsed |
| Power: | Max. 8 Watt at 50 Ohm |
RF Capsulotomy
| Frequency: | 500 kHz, pulsed |
| Power: | Max. 9.9. Watt at 50 Ohm, automatic regulation, 2 ranges |
Vitrectomy
| Control: | I/A/Cut or I/Cut/A | |||
| Drive: Memories: |
|
Power Requirements
| Electric: | 100-240 V, 50-60 Hz, power consumption 200 VA |
Dimensions and Weight
| Dimensions: | 200 x 241 x 164mm (without carrying handle) |
| Weight: | 5.6kg (pedal 2.7kg) |